Diagnosing deficiency of coagulation factor XII
Determining cause of prolonged activated partial thromboplastin time
Optical Clot-Based
Coagulation Factor XII Activity Assay, Plasma
Factor XII Activity Assay
Factor XII
Plasma Na Cit
Coagulation testing is highly complex, often requiring the performance of multiple assays and correlation with clinical information. For that reason, consider ordering a Coagulation Consultation.
If priority specimen, mark request form, give reason, and request a call-back.
Specimen Type: Platelet-poor plasma
Patient Preparation: Patient must not be receiving Coumadin (warfarin) or heparin therapy.
Collection Container/Tube: Light-blue top (3.2% sodium citrate)
Submission Container/Tube: Plastic vial
Specimen Volume: 1 mL
Collection Instructions:
1. Specimen must be collected prior to factor replacement therapy
2. For complete instructions, see Coagulation Guidelines for Specimen Handling and Processing.
3. Centrifuge, transfer all plasma into a plastic vial, and centrifuge plasma again.
4. Aliquot plasma into a plastic vial, leaving 0.25 mL in the bottom of centrifuged vial.
5. Freeze plasma immediately (no longer than 4 hours after collection) at -20 degrees C or ideally, at or below -40 degrees C.
Additional Information:
1. Double-centrifuged specimen is critical for accurate results as platelet contamination may cause spurious results.
2. Each coagulation assay requested should have its own vial.
If not ordering electronically, complete, print, and send a Coagulation Test Request (T753) with the specimen.
0.5 mL
Gross hemolysis | Reject |
Gross lipemia | Reject |
Gross icterus | Reject |
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Plasma Na Cit | Frozen | 14 days |
Diagnosing deficiency of coagulation factor XII
Determining cause of prolonged activated partial thromboplastin time
Factor XII is synthesized in the liver. Its biological half-life is 40 to 50 hours. Factor XII is a component of the contact activation system and is involved in both intrinsic pathway and fibrinolytic system.
Factor XII deficiency is often discovered when activated partial thromboplastin time is found to be unexpectedly long. The deficiency does not cause a known bleeding disorder.
An association between severe factor XII deficiency and thrombosis risk has been proposed but not proven.
Adults: 55-180%
Normal, full-term newborn infants or healthy premature infants may have decreased levels (> or =15% to 20%), which may not reach adult levels for 180 or more days postnatal.*
*See Pediatric Hemostasis References section in Coagulation Guidelines for Specimen Handling and Processing.
Acquired deficiency is associated with liver disease, nephritic syndrome, and chronic granulocytic leukemia.
Congenital homozygous deficiency: 20% activity
Congenital heterozygous deficiency: 20% to 50% activity
Deficiencies of other contact activator proteins (prekallikrein,
1. Renne T, Schmaier AH, Nickel KF, Blomback M, Maas C. In vivo roles of factor XII. Blood. 2012;120(22):4296-4303
2. Favaloro EJ, Lippi G, eds. Hemostasis and Thrombosis: Methods and Protocols. Humana Press; 2017
The factor XII assay is performed on the Instrumentation Laboratory ACL TOP using the activated partial thromboplastin time (aPTT) method and a factor-deficient substrate. Patient plasma is combined and incubated with a factor XII-deficient substrate (normal plasma depleted of factor XII by immunoadsorption) and an aPTT reagent. After a specified incubation time, calcium is added to trigger the coagulation process in the mixture. Then the time to clot formation is measured optically at a wavelength of 671 nm.(Owen CA Jr, Bowie EJW, Thompson JH Jr. Diagnosis of Bleeding Disorders. 2nd ed. Little, Brown and Company; 1975; Cielsa B. Defects of plasma clotting factors. In: Hematology in Practice. 3rd ed. FA Davis; 2019:chap 17)
Monday through Saturday
This test has been modified from the manufacturer's instructions. Its performance characteristics were determined by Mayo Clinic in a manner consistent with CLIA requirements. This test has not been cleared or approved by the US Food and Drug Administration.
85280
Test Id | Test Order Name | Order LOINC Value |
---|---|---|
F_12 | Coag Factor XII Assay, P | 3232-6 |
Result Id | Test Result Name |
Result LOINC Value
Applies only to results expressed in units of measure originally reported by the performing laboratory. These values do not apply to results that are converted to other units of measure.
|
---|---|---|
F_12 | Coag Factor XII Assay, P | 3232-6 |