Test Catalog

Test Id : ARBI

Acetylcholine Receptor (Muscle AChR) Binding Antibody, Serum

Useful For
Suggests clinical disorders or settings where the test may be helpful

Supporting the diagnosis of autoimmune myasthenia gravis (MG) in adults and children

 

Distinguishing autoimmune from congenital MG in adults and children or other acquired forms of neuromuscular junction transmission disorders

 

An adjunct to the test for P/Q-type calcium channel binding antibodies as a diagnostic aid for Lambert-Eaton myasthenic syndrome

Testing Algorithm
Delineates situations when tests are added to the initial order. This includes reflex and additional tests.

This is the primary diagnostic test for myasthenia gravis.

Method Name
A short description of the method used to perform the test

Radioimmunoassay (RIA)

NY State Available
Indicates the status of NY State approval and if the test is orderable for NY State clients.

Yes

Reporting Name
Lists a shorter or abbreviated version of the Published Name for a test

ACh Receptor (Muscle) Binding Ab

Aliases
Lists additional common names for a test, as an aid in searching

Acetylcholine Receptor (Muscle AChR) Antibodies

AChR (Acetylcholine Receptor)

Anti -Neuromuscular Junction Receptor Antibodies

Myasthenia Gravis Antibodies

Testing Algorithm
Delineates situations when tests are added to the initial order. This includes reflex and additional tests.

This is the primary diagnostic test for myasthenia gravis.

Specimen Type
Describes the specimen type validated for testing

Serum

Ordering Guidance

Standalone testing (this test) is recommended in certain situations.

 

This test should not be requested in patients who have recently received radioisotopes, therapeutically or diagnostically, because of potential assay interference. The specific waiting period before specimen collection will depend on the isotope administered, the dose given, and the clearance rate in the individual patient. Specimens will be screened for radioactivity prior to analysis. Radioactive specimens received in the laboratory will be held for 1 week and assayed if sufficiently decayed or canceled if radioactivity remains.

Specimen Required
Defines the optimal specimen required to perform the test and the preferred volume to complete testing

Patient Preparation: For optimal antibody detection, specimen collection is recommended prior to initiation of immunosuppressant medication.

Supplies: Sarstedt Aliquot Tube 5 mL (T914)

Collection Container/Tube:

Preferred: Red top

Acceptable: Serum gel

Submission Container/Tube: Plastic vial

Specimen Volume: 1.5 mL

Forms

If not ordering electronically, complete, print, and send a Neurology Specialty Testing Client Test Request (T732) with the specimen.

Specimen Minimum Volume
Defines the amount of sample necessary to provide a clinically relevant result as determined by the testing laboratory. The minimum volume is sufficient for one attempt at testing.

1 mL

Reject Due To
Identifies specimen types and conditions that may cause the specimen to be rejected

Gross hemolysis Reject
Gross lipemia Reject
Gross icterus Reject

Specimen Stability Information
Provides a description of the temperatures required to transport a specimen to the performing laboratory, alternate acceptable temperatures are also included

Specimen Type Temperature Time Special Container
Serum Refrigerated (preferred) 28 days
Frozen 28 days
Ambient 72 hours

Useful For
Suggests clinical disorders or settings where the test may be helpful

Supporting the diagnosis of autoimmune myasthenia gravis (MG) in adults and children

 

Distinguishing autoimmune from congenital MG in adults and children or other acquired forms of neuromuscular junction transmission disorders

 

An adjunct to the test for P/Q-type calcium channel binding antibodies as a diagnostic aid for Lambert-Eaton myasthenic syndrome

Testing Algorithm
Delineates situations when tests are added to the initial order. This includes reflex and additional tests.

This is the primary diagnostic test for myasthenia gravis.

Clinical Information
Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test

Fatigable weakness due to impaired postsynaptic transmission at the neuromuscular junction is characteristic of myasthenia gravis (MG). A clinical diagnosis should be supported by electrodiagnostic testing (ie, clinical-electrodiagnosis [EDX]). Positive autoimmune serology increases certainty of MG diagnosis but needs to be interpreted in the proper clinical-EDX context with response to anticholinesterase medications supporting the diagnosis. Most cases are autoimmune and are caused by IgG autoantibodies binding to critical postsynaptic membrane molecules (nicotinic muscle acetylcholine receptor [AChR] or its interacting proteins, such as muscle-specific kinase). Serologically, the detection of AChR binding antibody provides the best diagnostic sensitivity. However, the presence of both AChR binding and modulating activity improves diagnostic accuracy. Autoantibody detection frequency is lowest in patients with weakness confined to extraocular muscles (approximately 70% are positive for AChR binding antibodies) and highest in patients with generalized weakness due to MG (approximately 90% are positive for AChR binding antibodies). In adults with MG and AChR antibodies, approximately 20% will have thymoma and, very rarely (<1%), extrathymic cancers. Computed tomography imaging of the chest is considered the standard of care to evaluate for thymoma.

 

These results should only be interpreted in the appropriate clinical and electrophysiological context and are not diagnostic in isolation.

 

Note: Single antibody tests may be requested in the follow-up of patients with positive results previously documented in this laboratory.

Reference Values
Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.

< or =0.02 nmol/L

Interpretation
Provides information to assist in interpretation of the test results

Positive results (>0.02 nmol/L) are indicative of autoimmune myasthenia gravis (MG). These results should be interpreted in the appropriate clinical and electrophysiological context.

 

With a diagnosis of MG, a paraneoplastic basis should be considered with thymoma being the most frequently associated tumor with MG.

 

The clinical sensitivity of this assay is approximately 90% in nonimmunosuppressed patients with generalized MG. The frequency of antibody detection is lower in MG patients with weakness clinically restricted to ocular muscles (71%), and antibody titers are generally low in ocular MG (eg, 0.03-1.0 nmol/L).

 

Negative results do not exclude the diagnosis of MG. If clinical suspicion remains and symptoms persist or worsen consider retesting. Results may be negative in the first 12 months after symptoms of MG appear or during immunosuppressant therapy. Note: In follow up of seronegative patients with adult-acquired generalized MG, 17.4% seroconvert to positive at 12 months (ie, seronegativity rate at 12 months is 8.4%). A subset of MG patients that are persistently negative for acetylcholine receptor binding antibodies will have muscle-specific kinase (MuSK) antibodies, and therefore, it is recommended to test for MuSK antibodies in seronegative patients with high clinical suspicion of MG.

 

In general, there is not a close correlation between antibody titer and severity of weakness, but in individual patients, clinical improvement may be accompanied by a decrease in titer.

Cautions
Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances

The presence of elevated immunoglobulins due to therapeutic intervention or other disorders (ie, hypergammaglobulinemia) may lead to false-positive results.

 

Positive results may be found in some patients with Lambert-Eaton syndrome, paraneoplastic central nervous system, and peripheral nervous system autoimmune disorders and in healthy individuals.

 

The presence of alpha-bungarotoxin antibodies may interfere with this assay.

 

Specimens ideally should be collected prior to initiation of immunosuppressive therapies as these may reduce the sensitivity of this test.

Clinical Reference
Recommendations for in-depth reading of a clinical nature

1. Lennon VA. Serological profile of myasthenia gravis and distinction from the Lambert-Eaton myasthenic syndrome. Neurology. 1997;48(Suppl 5):S23-S27. doi:10.1212/WNL.48.Suppl_5.23S

2. Lachance DH, Lennon VA. Paraneoplastic neurological autoimmunity. In: Kalman B, Brannagan III T, eds. Neuroimmunology in Clinical Practice. Blackwell Publishing Ltd; 2008:210-217

3. Gilhus NE. Myasthenia gravis. N Engl J Med. 2016;375(26):2570-2581. doi:10.1056/NEJMra1602678

4. Nicolle MW. Myasthenia gravis and Lambert-Eaton myasthenic syndrome. Continuum (Minneap Minn). 2016;22(6, Muscle and Neuromuscular Junction Disorders):1978-2005. doi:10.1212/CON.0000000000000415

5. Shelly S, Paul P, Bi H, et al. Improving accuracy of myasthenia gravis autoantibody testing by reflex algorithm. Neurology. 2020;95(22):e3002-e3011. doi:10.1212/WNL.0000000000010910

Method Description
Describes how the test is performed and provides a method-specific reference

Fetal and adult detergent-solubilized, acetylcholine receptors (extracted from cultures of rhabdomyosarcoma [RD] cells) labeled with (125)I-alpha-bungarotoxin are incubated with patient sample. Anti-human IgG is then added to form an immunoprecipitate. After washing the precipitated immune complexes, the amount of (125)I-labeled receptor in the immunoprecipitate is measured using a gamma-counter. The amount of gamma emission in the precipitate is proportional to the amount of AChR-IgG in the sample. Results are reported in units of precipitated antigen (nMol) per L of patient sample.(Griesmann GE, Kryzer TJ, Lennon VA. Autoantibody profiles of myasthenia gravis and Lambert-Eaton myasthenic syndrome. In: Rose NR, Hamilton RG, eds. Manual of Clinical and Laboratory Immunology. 6th ed. ASM Press; 2002:1005-1012; Waters P, Pettingill P, Lang B. Detection methods for neural autoantibodies. Handb Clin Neurol. 2016;133:147-163. doi:10.1016/B978-0-444-63432-0.00009-8)

PDF Report
Indicates whether the report includes an additional document with charts, images or other enriched information

No

Day(s) Performed
Outlines the days the test is performed. This field reflects the day that the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time before the test is performed. Some tests are listed as continuously performed, which means that assays are performed multiple times during the day.

Monday through Sunday

Report Available
The interval of time (receipt of sample at Mayo Clinic Laboratories to results available) taking into account standard setup days and weekends. The first day is the time that it typically takes for a result to be available. The last day is the time it might take, accounting for any necessary repeated testing.

3 to 6 days

Specimen Retention Time
Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded

28 days

Performing Laboratory Location
Indicates the location of the laboratory that performs the test

Rochester

Fees :
Several factors determine the fee charged to perform a test. Contact your U.S. or International Regional Manager for information about establishing a fee schedule or to learn more about resources to optimize test selection.

  • Authorized users can sign in to Test Prices for detailed fee information.
  • Clients without access to Test Prices can contact Customer Service 24 hours a day, seven days a week.
  • Prospective clients should contact their account representative. For assistance, contact Customer Service.

Test Classification
Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR) product.

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information
Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Clinic Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.

CPT codes are provided by the performing laboratory.

86041

LOINC® Information
Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the order and results codes of this test. LOINC values are provided by the performing laboratory.

Test Id Test Order Name Order LOINC Value
ARBI ACh Receptor (Muscle) Binding Ab 97558-1
Result Id Test Result Name Result LOINC Value
Applies only to results expressed in units of measure originally reported by the performing laboratory. These values do not apply to results that are converted to other units of measure.
8338 ACh Receptor (Muscle) Binding Ab 97558-1

Test Setup Resources

Setup Files
Test setup information contains test file definition details to support order and result interfacing between Mayo Clinic Laboratories and your Laboratory Information System.

Excel | Pdf

Sample Reports
Normal and Abnormal sample reports are provided as references for report appearance.

Normal Reports | Abnormal Reports

SI Sample Reports
International System (SI) of Unit reports are provided for a limited number of tests. These reports are intended for international account use and are only available through MayoLINK accounts that have been defined to receive them.

SI Normal Reports | SI Abnormal Reports

Test Update Resources

Change Type Effective Date
Test Status - Test Resumed 2023-09-06
Test Status - Test Down 2023-08-09
Test Status - Test Delay 2023-07-20