Detecting drug abuse involving, amphetamines, barbiturates, benzodiazepines, cocaine, ethanol, methadone, opiates, phencyclidine, and tetrahydrocannabinol
This test is intended to be used in a setting where the test results can be used definitively to make a diagnosis.
This test is not intended for use in employment-related testing.
Test Id | Reporting Name | Available Separately | Always Performed |
---|---|---|---|
OPATU | Opiate Confirmation, U | Yes | No |
AMPHU | Amphetamines Confirmation, U | Yes | No |
BARBU | Barbiturates Confirmation, U | Yes | No |
COKEU | Cocaine and metabolite Conf, U | Yes | No |
ETOH | Ethanol, U | No | No |
MTDNU | Methadone Confirmation, U | Yes | No |
PCPU | Phencyclidine Confirmation, U | Yes | No |
THCU | Carboxy-THC Confirmation, U | Yes | No |
BNZU | Benzodiazepines Confirmation, U | Yes | No |
Test Id | Reporting Name | Available Separately | Always Performed |
---|---|---|---|
COCH | Chain of Custody Processing | No | Yes |
ADLTX | Adulterants Survey, CoC, U | Yes | Yes |
Testing begins with screening tests for alcohol and drugs of abuse. Positives are confirmed and quantitated by definitive methods (gas chromatography with flame ionization detector for ethanol; gas chromatography mass spectrometry for barbiturates, cocaine and metabolites, methadone, and phencyclidine) at an additional charge. Amphetamines, benzodiazepines, opiates, and tetrahydrocannabinol metabolite that screen positive will be quantified with liquid chromatography tandem mass spectrometry at an additional charge.
Enzymatic Assay/Immunoassay
ETOH: Headspace Gas Chromatography with Flame Ionization Detector (HSGC-FID)
7-Amino Flunitrazepam (Metabolite of Flunitrazepam) (Rohypnol)
Alcohol
Alprazolam (Xanax)
Amobarbital (Amytal)
Amphetamines
Amytal (Amobarbital)
Angel Dust (Phencyclidine)
Ativan (Lorazepam)
Barbita (Phenobarbital)
Barbiturates
Benzodiazepines
Benzoylecgonine (Cocaine Metabolite)
Butabarbital (Butisol)
Butalbital (Fiorinal)
Butisol (Butabarbital)
Cannabinoids (Tetrahydrocannabinol)
Chlordiazepoxide (Librium)
Clorazepate (Tranxene)
Cocaine
Codeine
Coke (Cocaine)
Crack (Cocaine)
Dalmane (Flurazepam)
Date Rape Drug (Rohypnol [Flunitrazepam])
Desalkyl Flurazepam (Metabolite)
Desoxyn (Methamphetamines)
Diazepam (Valium)
Dilaudid (Hydromorphone)
Dolophine (Methadone)
Drugs of Abuse
Ecstasy
Ethanol
Ethyl Alcohol
ETOH (Ethanol)
Fiorinal (Butalbital)
Flunitrazepam (Rohypnol)
Flurazepam (Dalmane)
Halicon (Triazolam)
Heroin
Hycodan (Hydrocodone)
Hydrocodone (Hycodan, Vicodin)
Hydromorphone (Dilaudid, Vicodin)
Hydroxy-Ethyl Flurazepam (Metabolite of Flurazepam) (Dalmane)
Killer Weed (Phencyclidine)
Librium (Chlordiazepoxide)
Lorazepam (Ativan)
Lortab (Hydromorphone)
Luminal (Phenobarbital)
Marijuana (Tetrahydrocannabinol)
MDA (Methylenedioxyamphetamine) Metabolite for Methylenedioxyethylamphetamine (MDEA)
MDMA (Methylenedioxymethamphetamine)
Mebaral (Mephobarbital)
Mephobarbital (Mebaral)
Methadone (Dolophine)
Methamphetamines (Desoxyn)
Methylenedioxyamphetamine (MDA) Metabolite for Methylenedioxyethylamphetamine (MDEA)
Methylenedioxymethamphetamine (MDMA)
Morphine
Nembutal (Pentobarbital)
Nordizepam (Tranxene)
Opiates
Oxazepam (Serax)
Oxycodone (Oxycontin, Percodan)
Oxycontin (Oxycodone)
PCP (Phencyclidine)
Pentobarbital (Nembutal)
Pentothal (Thiopental)
Percodan (Oxycodone)
Phencyclidine (PCP)
Phenobarbital
Restoril (Temazepam)
Rocket Fuel (Phencyclidine)
Rohypnol (Flunitrazepam)
Secobarbital (Seconal)
Seconal (Secobarbital)
Serax (Oxazepam)
Solfoton (Phenobarbital)
Speed (Amphetamines)
TCP (Phencyclidine)
Temazepam (Restoril)
Tetrahydrocannabinol (THC)
THC (Tetrahydrocannabinol)
Thiopental (Pentothal)
Toxicology Screen, Drugs
Tranxene (Clorazepate)
Triazolam (Halcion)
Tuinal (Amobarbital and Secobarbital)
Valium (Diazepam)
Vicodin (Hydrocodone)
Xanax (Alprazolam)
EDDP
Methadone metabolite
Testing begins with screening tests for alcohol and drugs of abuse. Positives are confirmed and quantitated by definitive methods (gas chromatography with flame ionization detector for ethanol; gas chromatography mass spectrometry for barbiturates, cocaine and metabolites, methadone, and phencyclidine) at an additional charge. Amphetamines, benzodiazepines, opiates, and tetrahydrocannabinol metabolite that screen positive will be quantified with liquid chromatography tandem mass spectrometry at an additional charge.
Urine
For situations where chain-of-custody is required, a Chain-of-Custody Kit (T282) is available. For chain-of-custody testing, order CDA7X / Drug Abuse Survey with Confirmation, Panel 9, Chain of Custody, Random, Urine.
If urine creatinine is required or adulteration of the sample is suspected, the following test should be requested, ADULT / Adulterants Survey, Urine. For additional information, please refer to ADULT / Adulterants Survey, Random, Urine.
Supplies: Urine Container, 60 mL (T313)
Collection Container/Tube: Plastic urine container
Submission Container/Tube: Plastic, 60-mL urine bottle
Specimen Volume: 30 mL
Collection Instructions:
1. Collect a random urine specimen.
2. Submit 30 mL in 1 plastic bottle.
3. No preservative.
Additional Information:
1. No specimen substitutions.
2. Additional drug panels and specific requests are available. Call 800-533-1710 or 507-266-5700.
3. Submitting less than 30 mL will compromise our ability to perform all necessary testing.
If not ordering electronically, complete, print, and send a Therapeutics Test Request (T831) with the specimen.
15 mL
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Urine | Refrigerated (preferred) | 7 days | |
Frozen | 14 days |
Detecting drug abuse involving, amphetamines, barbiturates, benzodiazepines, cocaine, ethanol, methadone, opiates, phencyclidine, and tetrahydrocannabinol
This test is intended to be used in a setting where the test results can be used definitively to make a diagnosis.
This test is not intended for use in employment-related testing.
Testing begins with screening tests for alcohol and drugs of abuse. Positives are confirmed and quantitated by definitive methods (gas chromatography with flame ionization detector for ethanol; gas chromatography mass spectrometry for barbiturates, cocaine and metabolites, methadone, and phencyclidine) at an additional charge. Amphetamines, benzodiazepines, opiates, and tetrahydrocannabinol metabolite that screen positive will be quantified with liquid chromatography tandem mass spectrometry at an additional charge.
This assay was designed to screen by immunoassay and confirm the following drugs, by either gas chromatography-mass spectrometry (GC-MS), gas chromatography-flame ionization detection (GC-FID), or liquid chromatography-tandem mass spectrometry (LC-MS/MS):
-Amphetamines
-Barbiturates
-Benzodiazepines
-Cocaine
-Ethanol
-Opiates
-Methadone
-Phencyclidine
-Tetrahydrocannabinol
This test uses the simple screening technique which involves immunoassay testing for drugs by class. All positive immunoassay screening results will be confirmed by the definitive assay available and is described in each individual reflex test (eg, AMPHU / Amphetamines Confirmation, Random, Urine). All positive screening results are confirmed by either GC-MS, GC-FID, or LC-MS/MS and quantitated before a positive result is reported.
Negative
Screening cutoff concentrations
Amphetamines: 500 ng/mL
Barbiturates: 200 ng/mL
Benzodiazepines: 100 ng/mL
Cocaine (benzoylecgonine-cocaine metabolite): 150 ng/mL
Ethanol: 10 mg/dL
Methadone metabolite: 300 ng/mL
Opiates: 300 ng/mL
Phencyclidine: 25 ng/mL
Tetrahydrocannabinol carboxylic acid: 50 ng/mL
This report is intended for use in clinical monitoring or management of patients. It is not intended for use in employment-related testing.
A positive result indicates that the patient has used the drugs detected in the recent past. See individual tests (eg, AMPHU / Amphetamines Confirmation, Random, Urine) for more information.
For information about drug testing, including estimated detection times, see Drugs of Abuse Testing.
The test does not screen for drug classes other than those listed above. More comprehensive screening is available using the serum or urine drug screens (DSS / Drug Screen, Prescription/Over the Counter, Serum or PDSU / Drug Screen, Prescription/Over the Counter, Random, Urine).
1. Physician's Desk Reference (PDR). 60th edition. Medical Economics Company; 2006
2. Bruntman LL. Goodman and Gilman's The Pharmacological Basis of Therapeutics. 11th ed. McGraw-Hill Book Company; 2006
3. Langman LJ, Bechtel LK, Holstege CP. Clinical toxicology. In: Rifai N, Chiu RWK, Young I, Burnham CAD, Wittwer CT, eds. Tietz Textbook of Laboratory Medicine. 7th ed. Elsevier; 2023: chap 43
The amphetamines, barbiturates, benzodiazepines, cocaine, methadone metabolite, opiates, phencyclidine, and tetrahydrocannabinol metabolite assays are based on the kinetic interaction of microparticles in a solution (KIMS) as measured by changes in light transmission. In the absence of sample drug, soluble drug conjugates bind to antibody-bound microparticles, causing the formation of particle aggregates. As the aggregation reaction proceeds in the absence of sample drug, the absorbance increases. When a urine sample contains the drug in question, this drug competes with the drug derivative conjugate for microparticle-bound antibody. Antibody bound to sample drug is no longer available to promote particle aggregation, and subsequent particle lattice formation is inhibited. The presence of sample drug diminishes the increasing absorbance in proportion to the concentration of drug in the sample. Sample drug content is determined relative to the value obtained for a known cutoff concentration of drug.(Package inserts: EDDP Specific Urine Enzyme Immunoassay. Immunalysis; 09/2018; AMPS2 cobas. Roche Diagnostics; V 10.0 09/2018; BARB cobas. Roche Diagnostics; V 13.0 09/2021; THC2 cobas. Roche Diagnostics; V 13.0 03/2022; BNZ2 cobas. Roche Diagnostics; V 2.0 04/2024; COC2 cobas. Roche Diagnostics; V 9.0 03/2019; ETOH2 cobas. Roche Diagnostics; V 16.0 06/2022; OPI2 cobas. Roche Diagnostics; V 16.0 01/2022; PCP cobas. Roche Diagnostics; V 13.0 09/2021)
ETOH Confirmation
Specimens are analyzed and quantified by headspace gas chromatography with flame ionization detection.(Baselt RC. Disposition of Toxic Drugs and Chemicals in Man, 10th edition, Biomedical Publications; 2014:2211)
Monday through Saturday
This test has been modified from the manufacturer's instructions. Its performance characteristics were determined by Mayo Clinic in a manner consistent with CLIA requirements. This test has not been cleared or approved by the US Food and Drug Administration.
80307
Test Id | Test Order Name | Order LOINC Value |
---|---|---|
CDAU7 | Confirmed Drug Abuse Panel 9, U | 87428-9 |
Result Id | Test Result Name |
Result LOINC Value
Applies only to results expressed in units of measure originally reported by the performing laboratory. These values do not apply to results that are converted to other units of measure.
|
---|---|---|
30909 | Alcohol | 34180-0 |
2573 | Amphetamines | 43983-6 |
2574 | Barbiturates | 70155-7 |
2575 | Benzodiazepines | 16195-0 |
21652 | Cocaine | 19359-9 |
4435 | Methadone metabolite | 16246-1 |
2577 | Opiates | 18390-5 |
2578 | Phencyclidine | 18392-1 |
2664 | Tetrahydrocannabinol | 19415-9 |
20672 | Chain of Custody | 77202-0 |