Monitoring trimipramine concentration during therapy
Evaluating potential trimipramine toxicity
May aid in evaluating patient compliance
Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS)
Surmontil (Trimipramine), Serum
Trimipramine, Serum
Surmontil (Trimipramine)
TCA (Tricyclic Antidepressants)
Tricyclic Antidepressants (TCA)
Serum Red
Supplies: Sarstedt Aliquot Tube, 5 mL (T914)
Collection Container/Tube: Red top (Serum gel/SST are not acceptable)
Submission Container/Tube: Plastic vial
Specimen Volume: 1 mL
Collection Instructions:
1. Collect specimen immediately before next scheduled dose (minimum 12 hours after last dose).
2. Centrifuge and aliquot serum into a plastic vial. Serum must be separated from cells within 2 hours of collection.
If not ordering electronically, complete, print, and send a Therapeutics Test Request (T831) with the specimen.
0.25 mL
Gross hemolysis | Reject |
Gross lipemia | Reject |
Gross icterus | Reject |
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum Red | Refrigerated (preferred) | 28 days | |
Frozen | 28 days | ||
Ambient | 7 days |
Monitoring trimipramine concentration during therapy
Evaluating potential trimipramine toxicity
May aid in evaluating patient compliance
Trimipramine is a tricyclic antidepressant with additional anxiety-reducing sedative activity. Daily dosages for adults range from 50 mg to 300 mg and are usually divided into 2 to 3 doses per day. Therapeutic ranges are based on serum samples collected at trough (ie, immediately before the next dose). Peak serum concentrations are typically achieved after 1 to 6 hours post dosage.
Common adverse effects include hypotension, tachycardia, constipation, dizziness, somnolence, and blurred vision. Risk of toxicity increases when concentrations exceed 500 ng/mL. Serious adverse effects include coma, seizures, and QRS prolongation with ventricular dysrhythmias.
Therapeutic concentration: 150-300 ng/mL
Note: Therapeutic ranges are for specimens collected at trough (ie, immediately before next scheduled dose). Levels may be elevated in non-trough specimens.
Most individuals display optimal response to trimipramine with serum levels of 150 to 300 ng/mL. Risk of toxicity is increased with trimipramine levels above 500 ng/mL.
Some individuals may respond well outside of this range or may display toxicity within the therapeutic range; thus, interpretation should include clinical evaluation.
Therapeutic ranges are based on specimens collected at trough (ie, immediately before the next dose).
This test cannot be performed on whole blood. Serum must be separated from cells within 2 hours of collection; if serum is not removed within this time, tricyclic antidepressant levels may be falsely elevated due to drug release from red blood cells.
Specimens that are obtained from gel tubes are not acceptable because the drug can absorb on the gel and lead to falsely decreased concentrations.
Coadministration of fluvoxamine, moclobemide, or quinidine inhibits the metabolism and markedly increases the serum concentrations of trimipramine.
1. Wille SM, Cooreman SG, Neels HM, Lambert WE. Relevant issues in the monitoring and the toxicology of antidepressants. Crit Rev Clin Lab Sci. 2008;45(1):25-89
2. Thanacoody HK, Thomas SHL. Antidepressant poisoning. Clin Med (Lond). 2003;3(2):114-118
3. Hiemke C, Bergemann N, Clement HW, et al. Consensus guidelines for therapeutic drug monitoring in neuropsychopharmacology: Update 2017. Pharmacopsychiatry. 2018;51(1-01):9-62
4. Milone MC, Shaw LM. Therapeutic drugs and their management. In: Rifai N, Chiu RWK, Young I, Burnham CAD, Wittwer CT, eds. Tietz Textbook of Laboratory Medicine. 7th ed. Elsevier; 2023:420-453
The tricyclic antidepressants are extracted from serum using a solvent to precipitate proteins. The supernatant is removed, and analysis is by liquid chromatography tandem mass spectrometry.(Unpublished Mayo method)
Monday, Wednesday, Friday
This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.
80299
Test Id | Test Order Name | Order LOINC Value |
---|---|---|
TRMP | Trimipramine, S | 4083-2 |
Result Id | Test Result Name |
Result LOINC Value
Applies only to results expressed in units of measure originally reported by the performing laboratory. These values do not apply to results that are converted to other units of measure.
|
---|---|---|
64269 | Trimipramine, S | 4083-2 |