Test Id : CURC

Copper/Creatinine Ratio, Urine

Useful For
Suggests clinical disorders or settings where the test may be helpful

Measurement of copper concentration of a part of the investigation of Wilson disease and obstructive liver disease using a random urine specimen

Method Name
A short description of the method used to perform the test

Only orderable as part of a profile. For more information see CURCU / Copper/Creatinine Ratio, Random, Urine.

 

Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

NY State Available
Indicates the status of NY State approval and if the test is orderable for NY State clients.

Yes

Reporting Name
Lists a shorter or abbreviated version of the Published Name for a test

Copper/Creat Ratio, U

Specimen Type
Describes the specimen type validated for testing

Urine

Specimen Required
Defines the optimal specimen required to perform the test and the preferred volume to complete testing

Only orderable as part of a profile. For more information see CURCU / Copper/Creatinine Ratio, Random, Urine.

 

Patient Preparation: High concentrations of barium are known to interfere with most metal tests. If barium-containing contrast media has been administered, the specimen should not be collected for at least 96 hours.

Supplies: Urine Tubes, 10 mL (T068)

Collection Container/Tube: Clean, plastic urine collection container with no metal cap or glued insert

Submission Container/Tube: Plastic, 10-mL urine tube or a clean, plastic aliquot container with no metal cap or glued insert

Specimen Volume: 3 mL

Collection Instructions:

1. Collect a random urine specimen.

2. See Metals Analysis Specimen Collection and Transport for complete instructions.

Special Instructions
Library of PDFs including pertinent information and forms related to the test

Specimen Minimum Volume
Defines the amount of sample necessary to provide a clinically relevant result as determined by the testing laboratory. The minimum volume is sufficient for one attempt at testing.

2 mL

Reject Due To
Identifies specimen types and conditions that may cause the specimen to be rejected

  All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.

Specimen Stability Information
Provides a description of the temperatures required to transport a specimen to the performing laboratory, alternate acceptable temperatures are also included

Specimen Type Temperature Time Special Container
Urine Refrigerated (preferred) 28 days
Ambient 28 days
Frozen 28 days

Useful For
Suggests clinical disorders or settings where the test may be helpful

Measurement of copper concentration of a part of the investigation of Wilson disease and obstructive liver disease using a random urine specimen

Clinical Information
Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test

The biliary system is the major pathway of copper excretion. Biliary excretion of copper requires an adenosine triphosphate (ATP)-dependent transporter protein. Variants in the gene for the transporter protein cause hepatolenticular degeneration (Wilson disease). Ceruloplasmin, the primary copper-carrying protein in the blood, is also reduced in Wilson disease. Urine copper excretion is increased in Wilson disease due to a decreased serum binding of copper to ceruloplasmin or due to allelic variances in cellular metal ion transporters.

 

Hypercupriuria (increased urinary copper) is also found in hemochromatosis, biliary cirrhosis, thyrotoxicosis, various infections, and a variety of other acute, chronic, and malignant diseases (including leukemia). Urine copper concentrations are also elevated during pregnancy and in patients taking contraceptives or estrogens.

 

Low urine copper levels are seen in malnutrition, hypoproteinemia, malabsorption, and nephrotic syndrome. Increased zinc consumption interferes with normal copper absorption from the gastrointestinal tract causing hypocupremia.

Reference Values
Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.

Only orderable as part of a profile. For more information see CURCU / Copper/Creatinine Ratio, Random, Urine.

 

Not applicable

Interpretation
Provides information to assist in interpretation of the test results

Humans normally excrete less than 60 mcg/24 hour in the urine.

Urinary copper excretion greater than 60 mcg/24 hour may be seen in:

-Wilson disease

-Obstructive biliary disease (eg, primary biliary cirrhosis, primary sclerosing cholangitis)

-Nephrotic syndrome (due to leakage through the kidney)

-Chelation therapy

-Estrogen therapy

-Mega dosing of zinc-containing vitamins

Because ceruloplasmin is an acute phase reactant, urine copper is elevated during acute inflammation. During the recovery phase, urine copper is usually below normal, reflecting the expected physiologic response to replace the copper that was depleted during inflammation.

Cautions
Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances

No significant cautionary statements

Clinical Reference
Recommendations for in-depth reading of a clinical nature

1. Zorbas YG, Kakuris KK, Deogenov VA, Yerullis KB. Copper homeostasis during hypokinesia in healthy subjects with higher and lower copper consumption. Tr Elem Electro. 2008;25:169-178

2. Lech T, Sadlik JK: Contribution to the data on copper concentration in blood and urine in patients with Wilson's disease and in normal subjects. Biol Trace Elem Res. 2007;118(1):16-20

3. Czlonkowska A, Litwin T, Dusek P, et al. Wilson disease. Nat Rev Dis Primers. 2018;4(1):21. doi: 10.1038/s41572-018-0018-3

4. Rifai N, Horwath AR, Wittwer CT, eds. Tietz Textbook of Clinical Chemistry and Molecular Diagnostics. 6th ed. Elsevier; 2018

Method Description
Describes how the test is performed and provides a method-specific reference

The metal of interest is analyzed by inductively coupled plasma mass spectrometry.(Unpublished Mayo method)

PDF Report
Indicates whether the report includes an additional document with charts, images or other enriched information

No

Day(s) Performed
Outlines the days the test is performed. This field reflects the day that the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time before the test is performed. Some tests are listed as continuously performed, which means that assays are performed multiple times during the day.

Monday, Thursday

Report Available
The interval of time (receipt of sample at Mayo Clinic Laboratories to results available) taking into account standard setup days and weekends. The first day is the time that it typically takes for a result to be available. The last day is the time it might take, accounting for any necessary repeated testing.

2 to 5 days

Specimen Retention Time
Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded

14 days

Performing Laboratory Location
Indicates the location of the laboratory that performs the test

Rochester

Fees :
Several factors determine the fee charged to perform a test. Contact your U.S. or International Regional Manager for information about establishing a fee schedule or to learn more about resources to optimize test selection.

  • Authorized users can sign in to Test Prices for detailed fee information.
  • Clients without access to Test Prices can contact Customer Service 24 hours a day, seven days a week.
  • Prospective clients should contact their account representative. For assistance, contact Customer Service.

Test Classification
Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR) product.

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information
Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Clinic Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.

CPT codes are provided by the performing laboratory.

82525

LOINC® Information
Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the order and results codes of this test. LOINC values are provided by the performing laboratory.

Test Id Test Order Name Order LOINC Value
CURC Copper/Creat Ratio, U 13829-7
Result Id Test Result Name Result LOINC Value
Applies only to results expressed in units of measure originally reported by the performing laboratory. These values do not apply to results that are converted to other units of measure.
615258 Copper/Creat Ratio, U 13829-7

Test Setup Resources

Setup Files
Test setup information contains test file definition details to support order and result interfacing between Mayo Clinic Laboratories and your Laboratory Information System.

Excel | Pdf

Sample Reports
Normal and Abnormal sample reports are provided as references for report appearance.

Normal Reports | Abnormal Reports

SI Sample Reports
International System (SI) of Unit reports are provided for a limited number of tests. These reports are intended for international account use and are only available through MayoLINK accounts that have been defined to receive them.

SI Normal Reports | SI Abnormal Reports