Test Catalog

Test Id : F_2

Coagulation Factor II Activity Assay, Plasma

Useful For
Suggests clinical disorders or settings where the test may be helpful

Diagnosing a congenital deficiency (rare) of coagulation factor II

 

Evaluating acquired deficiencies associated with liver disease or vitamin K deficiency, oral anticoagulant therapy, and antibody-induced deficiencies (eg, in association with lupus-like anticoagulant)

 

Determining warfarin treatment stabilization in patients with nonspecific inhibitors (ie, lupus anticoagulant)

 

Determining degree of anticoagulation with warfarin to correlate with level of protein S

 

Investigation of prolonged prothrombin time or activated partial thromboplastin time

Method Name
A short description of the method used to perform the test

Optical Clot-Based

NY State Available
Indicates the status of NY State approval and if the test is orderable for NY State clients.

No

Reporting Name
Lists a shorter or abbreviated version of the Published Name for a test

Coag Factor II Assay, P

Aliases
Lists additional common names for a test, as an aid in searching

FACTOR II ACTIVITY

Factor II, Plasma

Prothrombin

Factor II

Specimen Type
Describes the specimen type validated for testing

Plasma Na Cit

Ordering Guidance

Coagulation testing is highly complex, often requiring the performance of multiple assays and correlation with clinical information. For that reason, we suggest ordering Coagulation Consultations.

Necessary Information

If priority specimen, mark request form, give reason, and request a call-back.

Specimen Required
Defines the optimal specimen required to perform the test and the preferred volume to complete testing

Specimen Type: Platelet-poor plasma

Patient Preparation:  Patient must not be receiving coumadin (warfarin) or heparin therapy. (If not possible for medical reasons, note on request.)

Collection Container/Tube: Light-blue top (3.2% sodium citrate)

Submission Container/Tube: Plastic vial

Specimen Volume: 1 mL

Collection Instructions:

1. Specimen must be collected prior to factor replacement therapy

2. For complete instructions, see Coagulation Guidelines for Specimen Handling and Processing

3. Centrifuge, transfer all plasma into a plastic vial, and centrifuge plasma again.

4. Aliquot plasma into a plastic vial, leaving 0.25 mL in the bottom of centrifuged vial.

5. Freeze plasma immediately (no longer than 4 hours after collection) at -20 degrees C or, ideally, -40 degrees C or below.

 

Additional Information:

1. Double-centrifuged specimen is critical for accurate results as platelet contamination may cause spurious results.

2. Each coagulation assay requested should have its own vial.

Special Instructions
Library of PDFs including pertinent information and forms related to the test

Forms

Specimen Minimum Volume
Defines the amount of sample necessary to provide a clinically relevant result as determined by the testing laboratory. The minimum volume is sufficient for one attempt at testing.

0.5 mL

Reject Due To
Identifies specimen types and conditions that may cause the specimen to be rejected

Gross hemolysis Reject
Gross lipemia Reject
Gross icterus Reject

Specimen Stability Information
Provides a description of the temperatures required to transport a specimen to the performing laboratory, alternate acceptable temperatures are also included

Specimen Type Temperature Time Special Container
Plasma Na Cit Frozen 14 days

Useful For
Suggests clinical disorders or settings where the test may be helpful

Diagnosing a congenital deficiency (rare) of coagulation factor II

 

Evaluating acquired deficiencies associated with liver disease or vitamin K deficiency, oral anticoagulant therapy, and antibody-induced deficiencies (eg, in association with lupus-like anticoagulant)

 

Determining warfarin treatment stabilization in patients with nonspecific inhibitors (ie, lupus anticoagulant)

 

Determining degree of anticoagulation with warfarin to correlate with level of protein S

 

Investigation of prolonged prothrombin time or activated partial thromboplastin time

Clinical Information
Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test

Factor II (prothrombin) is a vitamin K-dependent serine protease synthesized in liver. It participates in the final common pathway of coagulation, as the substrate for the prothrombinase enzyme complex. Prothrombin is the precursor of thrombin (IIa), which converts fibrinogen to fibrin. Plasma biological half-life is about 3 days.

 

Deficiency of factor II may cause prolonged prothrombin time and activated partial thromboplastin time. Deficiency may result in a bleeding diathesis.

Reference Values
Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.

Adults: 75-145%

Normal, full-term newborn infants or healthy premature infants may have decreased levels (> or =25%) which may remain below adult levels for > or =180 days postnatal.*

 

*See Pediatric Hemostasis References section in Coagulation Guidelines for Specimen Handling and Processing

Interpretation
Provides information to assist in interpretation of the test results

Liver disease, vitamin K deficiency, or warfarin anticoagulation can cause decreased factor II activity.

 

Normal newborn infants may have levels of 25% to 50%.

Cautions
Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances

Factor II is one of the last vitamin K-dependent coagulation factors to decrease after starting warfarin therapy and one of the last to return to normal when anticoagulation is discontinued. It may take 10 to 14 days for a return to baseline levels.

Clinical Reference
Recommendations for in-depth reading of a clinical nature

1. Lancellotti S, De Cristofaro R. Congenital prothrombin deficiency. Semin Thromb Hemost. 2009;35(4):367-381. doi:10.1055/s-0029-1225759

2. Peyvandi F, Bolton-Maggs PH, Batorova A, De Moerloose P. Rare bleeding disorders. Haemophilia. 2012;18 Suppl 4:148-153. doi:10.1111/j.1365-2516.2012.02841.x

3. Girolami A, Scandellari R, Scapin M, Vettore S. Congenital bleeding disorders of the vitamin K-dependent clotting factors. Vitam Horm. 2008;78:281-374. doi:10.1016/S0083-6729(07)00014-3

4. Brenner B, Kuperman AA, Watzka M, Oldenburg J. Vitamin K-dependent coagulation factors deficiency. Semin Thromb Hemost. 2009;35(4):439-446. doi:10.1055/s-0029-1225766

5. Favaloro EJ and Lippi G. eds. Hemostasis and Thrombosis, Methods and Protocols. Humana Press 2017

Method Description
Describes how the test is performed and provides a method-specific reference

The factor II assay is performed on the Instrument Laboratory ACL TOP using the prothrombin time (PT) method and a factor-deficient substrate. Patient plasma is combined and incubated with a factor II-deficient substrate (normal plasma depleted of factor II by immunoadsorption). After a specified incubation time, a PT reagent is added to trigger the coagulation process in the mixture. Then the time to clot formation is measured optically at a wavelength of 671 nm.(Owen CA Jr, Bowie EJW, Thompson JH Jr: Diagnosis of Bleeding Disorders. 2nd ed. Little, Brown and Company; 1975; Meijer P, Verbruggen HW, Spannagi M: Clotting factors and inhibitors: Assays and interpretation. In: Kottke-Marchant K, ed. Laboratory Hematology Practice. Wiley Blackwell Publishing; 2012:435-446)

PDF Report
Indicates whether the report includes an additional document with charts, images or other enriched information

No

Day(s) Performed
Outlines the days the test is performed. This field reflects the day that the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time before the test is performed. Some tests are listed as continuously performed, which means that assays are performed multiple times during the day.

Monday through Friday

Report Available
The interval of time (receipt of sample at Mayo Clinic Laboratories to results available) taking into account standard setup days and weekends. The first day is the time that it typically takes for a result to be available. The last day is the time it might take, accounting for any necessary repeated testing.

1 to 3 days

Specimen Retention Time
Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded

7 days

Performing Laboratory Location
Indicates the location of the laboratory that performs the test

Jacksonville

Fees :
Several factors determine the fee charged to perform a test. Contact your U.S. or International Regional Manager for information about establishing a fee schedule or to learn more about resources to optimize test selection.

  • Authorized users can sign in to Test Prices for detailed fee information.
  • Clients without access to Test Prices can contact Customer Service 24 hours a day, seven days a week.
  • Prospective clients should contact their account representative. For assistance, contact Customer Service.

Test Classification
Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR) product.

This test has been modified from the manufacturer's instructions. Its performance characteristics were determined by Mayo Clinic in a manner consistent with CLIA requirements. This test has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information
Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Clinic Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.

CPT codes are provided by the performing laboratory.

85210

LOINC® Information
Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the order and results codes of this test. LOINC values are provided by the performing laboratory.

Test Id Test Order Name Order LOINC Value
F_2 Coag Factor II Assay, P 3289-6
Result Id Test Result Name Result LOINC Value
Applies only to results expressed in units of measure originally reported by the performing laboratory. These values do not apply to results that are converted to other units of measure.
F_2 Coag Factor II Assay, P 3289-6

Test Setup Resources

Setup Files
Test setup information contains test file definition details to support order and result interfacing between Mayo Clinic Laboratories and your Laboratory Information System.

Excel | Pdf

Sample Reports
Normal and Abnormal sample reports are provided as references for report appearance.

Normal Reports | Abnormal Reports

SI Sample Reports
International System (SI) of Unit reports are provided for a limited number of tests. These reports are intended for international account use and are only available through MayoLINK accounts that have been defined to receive them.

SI Normal Reports | SI Abnormal Reports