Test Catalog

Test Id : INS

Insulin, Serum

Useful For
Suggests clinical disorders or settings where the test may be helpful

Diagnosing insulinoma, when used in conjunction with proinsulin and C-peptide measurements

 

Management of diabetes mellitus

Method Name
A short description of the method used to perform the test

Electrochemiluminescence Immunoassay (ECLIA)

NY State Available
Indicates the status of NY State approval and if the test is orderable for NY State clients.

No

Reporting Name
Lists a shorter or abbreviated version of the Published Name for a test

Insulin, S

Aliases
Lists additional common names for a test, as an aid in searching

Fasting Insulin

Immunoreactive Insulin

Insulin Assay

Insulin Level, Serum

Native Insulin

Specimen Type
Describes the specimen type validated for testing

Serum

Ordering Guidance

Patients on insulin therapy may develop anti-insulin antibodies. These antibodies may interfere in the assay system, causing inaccurate results. In such individuals, measurement of free insulin INSFT / Insulin, Free and Total, Serum should be performed.

Specimen Required
Defines the optimal specimen required to perform the test and the preferred volume to complete testing

Patient Preparation:

1. Patient should be fasting.

2. For 12 hours before specimen collection do not take multivitamins or dietary supplements containing biotin (vitamin B7), which is commonly found in hair, skin, and nail supplements and multivitamins.

Supplies: Sarstedt 5 mL Aliquot Tube (T914)

Collection Container/Tube:

Preferred: Serum gel

Acceptable: Red top

Submission Container/Tube: Plastic vial

Specimen Volume: 1 mL

Collection Instructions:

1. Avoid hemolysis

2. Label specimens with corresponding collection times.

3. Centrifuge and aliquot serum into plastic vial within 2 hours of collection.

Additional Information: If multiple specimens are drawn, send separate order for each specimen.

Specimen Minimum Volume
Defines the amount of sample necessary to provide a clinically relevant result as determined by the testing laboratory. The minimum volume is sufficient for one attempt at testing.

0.75 mL

Reject Due To
Identifies specimen types and conditions that may cause the specimen to be rejected

Gross hemolysis Reject
Gross lipemia OK
Autopsy specimen Reject

Specimen Stability Information
Provides a description of the temperatures required to transport a specimen to the performing laboratory, alternate acceptable temperatures are also included

Specimen Type Temperature Time Special Container
Serum Frozen (preferred) 180 days
Refrigerated 7 days

Useful For
Suggests clinical disorders or settings where the test may be helpful

Diagnosing insulinoma, when used in conjunction with proinsulin and C-peptide measurements

 

Management of diabetes mellitus

Clinical Information
Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test

Insulin is a hormone produced by the beta cells of the pancreas. It regulates the uptake and utilization of glucose and is also involved in protein synthesis and triglyceride storage.

 

Type 1 diabetes (insulin-dependent diabetes) is caused by insulin deficiency due to destruction of insulin-producing pancreatic islet (beta) cells. Type 2 diabetes (noninsulin-dependent diabetes) is characterized by resistance to the action of insulin (insulin resistance).

 

Insulin levels may be increased in patients with pancreatic beta cell tumors (insulinoma).

Reference Values
Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.

2.6-24.9 mcIU/mL

 

For SI unit Reference Values, see https://www.mayocliniclabs.com/order-tests/si-unit-conversion.html

Interpretation
Provides information to assist in interpretation of the test results

During prolonged fasting, when the patient's glucose level is reduced to <40 mg/dL, an elevated insulin level plus elevated levels of proinsulin and C-peptide suggest insulinoma.

 

Insulin levels generally decline in patients with type 1 diabetes mellitus.

 

In the early stage of type 2 diabetes, insulin levels are either normal or elevated. In the late stage of type 2 diabetes, insulin levels decline.

 

In normal individuals, insulin levels parallel blood glucose levels.

 

To compare insulin and C-peptide concentrations (ie, insulin to C-peptide ratio):

-Convert insulin to pmol/L: insulin concentration in mcIU/mL x 6.945 = insulin concentration in pmol/L.

-Convert C-peptide to pmol/L: C-peptide concentration in ng/mL x 331 = C-peptide concentration in pmol/L.

Cautions
Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances

Human antimouse antibodies (HAMA) may interfere with the assay.

 

This assay has 100% cross-reactivity with recombinant human insulin (Novolin R and Novolin N). It does not recognize other commonly used analogues of injectable insulin (ie, insulin lispro, insulin aspart, and insulin glargine).

Clinical Reference
Recommendations for in-depth reading of a clinical nature

1. Threatte GA, Henry JB: Carbohydrates. In: Henry JB, ed. Clinical Diagnosis and Management by Laboratory Methods. 19th ed. WB Saunders Company; 1996:194-207

2. Sacks DB: Diabetes mellitus. In: Rifai N, Horvath AR, Wittwer CT, eds. Tietz Textbook of Clinical Chemistry and Molecular Diagnostics 6th ed. Elsevier; 2018:1160-1200

Method Description
Describes how the test is performed and provides a method-specific reference

Insulin from the sample, a biotinylated monoclonal insulin-specific antibody, and a monoclonal insulin-specific antibody labeled with a ruthenium complex form a sandwich complex. After addition of streptavidin-coated microparticles, the complex becomes bound to the solid phase via interaction of biotin and streptavidin. The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission, which is measured by a photomultiplier. Results are determined via a calibration curve that is instrument specifically generated by 2-point calibration and a master curve provided via the cobas link.(Package insert: Elecsys Insulin. Roche Diagnostics; v3.0, 06/2020)

PDF Report
Indicates whether the report includes an additional document with charts, images or other enriched information

No

Day(s) Performed
Outlines the days the test is performed. This field reflects the day that the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time before the test is performed. Some tests are listed as continuously performed, which means that assays are performed multiple times during the day.

Report Available
The interval of time (receipt of sample at Mayo Clinic Laboratories to results available) taking into account standard setup days and weekends. The first day is the time that it typically takes for a result to be available. The last day is the time it might take, accounting for any necessary repeated testing.

1 to 3 days

Specimen Retention Time
Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded

7 days

Performing Laboratory Location
Indicates the location of the laboratory that performs the test

Jacksonville

Fees :
Several factors determine the fee charged to perform a test. Contact your U.S. or International Regional Manager for information about establishing a fee schedule or to learn more about resources to optimize test selection.

  • Authorized users can sign in to Test Prices for detailed fee information.
  • Clients without access to Test Prices can contact Customer Service 24 hours a day, seven days a week.
  • Prospective clients should contact their account representative. For assistance, contact Customer Service.

Test Classification
Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR) product.

This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.

CPT Code Information
Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Clinic Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.

CPT codes are provided by the performing laboratory.

83525

LOINC® Information
Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the order and results codes of this test. LOINC values are provided by the performing laboratory.

Test Id Test Order Name Order LOINC Value
INS Insulin, S 27873-9
Result Id Test Result Name Result LOINC Value
Applies only to results expressed in units of measure originally reported by the performing laboratory. These values do not apply to results that are converted to other units of measure.
INS Insulin, S 27873-9

Test Setup Resources

Setup Files
Test setup information contains test file definition details to support order and result interfacing between Mayo Clinic Laboratories and your Laboratory Information System.

Excel | Pdf

Sample Reports
Normal and Abnormal sample reports are provided as references for report appearance.

Normal Reports | Abnormal Reports

SI Sample Reports
International System (SI) of Unit reports are provided for a limited number of tests. These reports are intended for international account use and are only available through MayoLINK accounts that have been defined to receive them.

SI Normal Reports | SI Abnormal Reports