Test Catalog

Test Id : NFLP

Neurofilament Light Chain, Plasma

Useful For
Suggests clinical disorders or settings where the test may be helpful

Assessing neuronal damage related to various neurodegenerative diseases

Method Name
A short description of the method used to perform the test

Chemiluminescent Enzyme Immunoassay

NY State Available
Indicates the status of NY State approval and if the test is orderable for NY State clients.

Yes

Reporting Name
Lists a shorter or abbreviated version of the Published Name for a test

Neurofilament Light Chain, P

Aliases
Lists additional common names for a test, as an aid in searching

Alzheimer

Dementia

Neurofilament

NfL

Specimen Type
Describes the specimen type validated for testing

EDTA Plasma

Specimen Required
Defines the optimal specimen required to perform the test and the preferred volume to complete testing

Supplies: Sarstedt Aliquot Tube, 5 mL (T914)

Collection Container/Tube:

Preferred: Lavender top (EDTA)

Acceptable: None

Submission Container/Tube: Plastic screw-top vial

Specimen Volume: 0.6 mL

Collection Information: Centrifuge and aliquot plasma into a plastic vial. Do not submit specimen in original tube.

Specimen Minimum Volume
Defines the amount of sample necessary to provide a clinically relevant result as determined by the testing laboratory. The minimum volume is sufficient for one attempt at testing.

0.50 mL

Reject Due To
Identifies specimen types and conditions that may cause the specimen to be rejected

Gross hemolysis Reject
Gross lipemia Reject
Gross icterus OK

Specimen Stability Information
Provides a description of the temperatures required to transport a specimen to the performing laboratory, alternate acceptable temperatures are also included

Specimen Type Temperature Time Special Container
EDTA Plasma Ambient 7 days
Refrigerated (preferred) 14 days
Frozen 90 days

Useful For
Suggests clinical disorders or settings where the test may be helpful

Assessing neuronal damage related to various neurodegenerative diseases

Clinical Information
Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test

Neurofilaments (NF) are exclusively located in the neuronal cytoskeleton and are released to the interstitial fluid upon axonal injury or neurodegeneration. NF concentrations in cerebrospinal fluid (CSF) and blood have been shown to correlate with the extent of axonal damage or neurodegeneration in various neurodegenerative diseases. Of the family of NF proteins, neurofilament light chain (NfL) has gained the most interest as a candidate marker of neurodegeneration. During axonal damage, NfL is released into the CSF, and eventually into the blood where concentrations are 40-fold lower than in the CSF. Concentrations of NfL in plasma have been shown to be approximately 5% to 10% lower than those measured in serum.

 

Circulating NfL concentrations increase with age with at a rate approximately 2% to 3% per year of age in both male and female individuals. While the specific cause of this increase has not been elucidated, it is believed to be related to the aging process as well as to the development of subclinical ischemic events. NfL concentrations in blood (plasma or serum) reflect the extent of axonal damage, making them a generic marker of disease activity. Increases in NfL concentrations have been reported in individuals with traumatic brain injury, amyotrophic lateral sclerosis, multiple sclerosis, frontotemporal dementia, Alzheimer disease (AD), and other neurodegenerative diseases.

 

Plasma neurofilament light chain (NfL) is a non-specific marker of neuro-axonal injury showing promising associations with outcomes in several neurological conditions. In neurodegenerative diseases, NfL may also serve as a prognostic marker of disease progression and drug efficacy biomarker of experimental therapies. In a meta-analysis of AD, frontotemporal dementia, and amyotrophic lateral sclerosis, plasma NfL concentrations were elevated in patients compared to controls with utility in differentiating neurodegenerative conditions from non-neurodegenerative mimics. However, due to a lack of specificity to a particular neurodegenerative disease, its role as a diagnostic marker may be limited.

Reference Values
Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.

<2.5 years: < or =12.8 pg/mL

2.5 to 4 years: < or =11.8 pg/mL

5 to 9 years: < or =10.4 pg/mL

10 to 14 years: < or =8.8 pg/mL

15 to 19 years: < or =9.2 pg/mL

20 to 24 years: < or =10.4 pg/mL

25 to 29 years: < or =11.9 pg/mL

30 to 34 years: < or =13.5 pg/mL

35 to 39 years: < or =15.3 pg/mL

40 to 44 years: < or =17.3 pg/mL

45 to 49 years: < or =19.7 pg/mL

50 to 54 years: < or =22.4 pg/mL

55 to 59 years: < or =25.4 pg/mL

60 to 64 years: < or =28.8 pg/mL

65 to 69 years: < or =32.7 pg/mL

70 to 74 years: < or =37.1 pg/mL

75 to 79 years: < or =42.1 pg/mL

80 to 84 years: < or = 47.8 pg/mL

> or =85 years: < or =54.3 pg/mL

Interpretation
Provides information to assist in interpretation of the test results

Interpretation of plasma neurofilament light chain (NfL) concentrations depends on the clinical context.

 

Normal plasma NfL concentrations are generally consistent with the absence of neurodegeneration. In patients receiving therapy for multiple sclerosis (MS), normal or decreased NfL concentrations would suggest treatment response and a more favorable prognosis.

 

Increased plasma NfL concentrations are consistent with the presence of neurodegeneration. In patients with a known neurologic condition, elevated NfL or increased concentrations from an established, patient-specific baseline may indicate poorer prognosis and/or disease progression.

 

In multiple sclerosis, NfL is most valuable as a prognostic indicator for severity of disease, disease progression, and as an indicator of response to therapy. Baseline plasma NfL concentrations are a valuable contribution to the initial workup in patients with diagnosed or suspected MS and should be interpreted in the context of other clinical information. The Consortium of Multiple Sclerosis Centers recommends measurement of blood NfL at baseline and regular follow-up (3-6 months) for obtaining prognostic information and evaluating treatment response.(1) Elevated baseline or increasing NfL concentrations can predict multiple sclerosis relapses and other disease activity. The use of blood NfL in serial disease monitoring and treatment response has been evaluated in various prospective clinical trials. Reductions in NfL concentrations after different treatments tend to follow the hierarchy of treatment efficacy, with greatest reductions observed with the most intensive treatments. A study that included over 1000 patients with MS receiving various treatments, reported the largest reductions in plasma NfL concentrations following alemtuzumab treatment (54% reduction), and the smallest reduction with teriflunomide treatment (7%).(2)

 

The nonspecific increase of NfL in a number of neurodegenerative disorders reduces the utility of NfL for differentiation of Alzheimer Disease (AD) from other cause of dementia. Measuring NfL in the context of AD likely has limited clinical utility.

 

In amyotrophic lateral sclerosis (ALS), NfL concentrations have been suggested to be able to discriminate ALS from ALS-mimics. NfL concentrations at symptom onset may be prognostic of disease progression rate and may be used to stratify patients into groups with a similar prognosis in clinical trials. Blood NfL concentrations remain relatively stable throughout the disease. A longitudinal decline in NfL concentrations has been described with some ALS treatments. For example, the US Food and Drug Administration has approved Qalsody (tofersen) to treat patients with ALS associated with a genetic variant in the superoxide dismutase 1 (SOD1) gene. The approval was based on a reduction in plasma NfL concentrations at the end treatment compared to the placebo arm. For other ALS therapies such as riluzole, NfL concentrations have been reported not to be useful for monitoring treatment effects.

 

Parkinson disease (PD) patients with elevated NfL concentrations have been reported to have worse cognitive decline, brain cortical atrophy, and motor scores. Blood NfL concentrations in atypical forms of Parkinson disease are higher than in PD and may be used to help differentiate PD from atypical parkinsonian disorders such as progressive supranuclear palsy, corticobasal degeneration, and multiple system atrophy.

 

In frontotemporal dementia (FTD), NfL concentrations differ according to the underlying mutation - they are highest in people with the GRN mutation and lowest in people with the MAPT mutation. These levels rise in the presymptomatic stages of FTD, and the timing of preclinical increases differs with the underlying mutation.

 

In traumatic brain injury (TBI), blood NfL concentrations have been evaluated both in the context of mild TBI (mTBI) diagnosis (acute setting) and prognosis (outcome prediction). In the acute setting, the utility of NfL in identifying mTBI within 24 hours of an injury has been controversial, likely due to the different timepoints used in studies for evaluating NfL concentrations after the injury (ranging from 1-, 4-, 6-, 12-, and 24-hours post-injury). A recent 2022 review describes the findings of six different publications looking at the role of NfL in acute mTBI concluding that, although the clinical usefulness of blood NfL for acute diagnosis of mild TBI is uncertain, the biomarker shows promise for the prognosis of complications of mild TBI, neuroimaging findings and recovery when measured during the first days to weeks after injury.(3)

 

In hypoxic–ischemic brain injury, NfL is a promising prognostic marker after cardiac arrest. NfL concentrations increase within the first 24 hours after cardiac arrest and the increased concentrations persist for days to months. A recent meta-analysis showed that elevated NfL concentrations 48 hours after cardiac arrest predict poor neurological outcomes.(4) Several studies have shown that the prognostic value of blood NfL in this context is higher than that of other blood biomarkers routinely used for cardiac arrest prognosis including neuron-specific enolase, S100 and total-Tau.

Cautions
Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances

Increases in neurofilament light chain (NfL) are not disease specific. Results should only be used in conjunction with other clinical information when evaluating patients with neurodegeneration.

 

Higher concentrations of NfL may be found in persons with history of stroke, atrial fibrillation, myocardial infarction, chronic kidney disease, pregnancy, and diabetes.

 

Lower concentrations of NfL may be found in individuals with a body mass index of 30 or more.

 

Neurofilament light chain concentrations obtained with different methods may be different and cannot be used interchangeably.

 

All immunometric assays can, on rare occasions, be subject to a hooking effect at extremely high analyte concentrations (false-low results), heterophilic antibody interference (false-high results), or autoantibody interference (unpredictable effects). If the laboratory result does not fit the clinical picture, these possibilities should be considered.

 

In rare cases, some individuals can develop antibodies to mouse or other animal antibodies (often referred to as human anti-mouse antibodies [HAMA] or heterophile antibodies), which may cause interference in some immunoassays. Caution should be used in interpretation of results and the laboratory should be alerted if the result does not correlate with the clinical presentation.

Clinical Reference
Recommendations for in-depth reading of a clinical nature

1. Freedman MS, Gnanapavan S, Booth RA, et al. Guidance for use of neurofilament light chain as a cerebrospinal fluid and blood biomarker in multiple sclerosis management. EBioMedicine. 2024;101:104970. doi:10.1016/j.ebiom.2024.104970

2. Delcoigne B, Manouchehrinia A, Barro C, et al. Blood neurofilament light levels segregate treatment effects in multiple sclerosis. Neurology. 2020;94(11):e1201-e1212. doi:10.1212/WNL.0000000000009097

3. Farragher CD, Ku Y, Powers JE. The potential role of neurofilament light in mild traumatic brain injury diagnosis: a systematic review. Cureus. 2022;14(11):e31301

4. Wang SL, Li N, Feng SY, Li Y. Serum neurofilament light chain as a predictive marker of neurologic outcome after cardiac arrest: a meta-analysis. BMC Cardiovasc. Disord. 2023;23(1):193

5. Khalil M, Teunissen CE, Lehmann S, et al. Neurofilaments as biomarkers in neurological disorders - towards clinical application. Nat Rev Neurol. 2024;20(5):269-287. doi:10.1038/s41582-024-00955-x

6. Ashton NJ, Janelidze S, Al Khleifat A, et al. A multicentre validation study of the diagnostic value of plasma neurofilament light. Nat Commun. 2021;12(1):3400

7. Wendel EM, Bertolini A, Kousoulos L, et al. Serum neurofilament light-chain levels in children with monophasic myelin oligodendrocyte glycoprotein-associated disease, multiple sclerosis, and other acquired demyelinating syndrome. Mult Scler. 2022;28(10):1553-1561

8. Karantali E, Kazis D, McKenna J, et al. Neurofilament light chain in patients with a concussion or head impacts: a systematic review and meta-analysis. Eur J Trauma Emerg Surg. 2022;48(3):1555-1567

9. Bornhorst JA, Figdore D, Campbell MR, et al. Plasma neurofilament light chain (NfL) reference interval determination in an age-stratified cognitively unimpaired cohort. Clin Chim Acta. 2022;535:153-156. doi:10.1016/j.cca.2022.08.017

10. Ashrafzadeh-Kian S, Figdore D, Larson B, et al. Head-to-head comparison of four plasma neurofilament light chain (NfL) immunoassays. Clin Chim Acta. 2024;561:119817. doi: 10.1016/j.cca.2024.119817

11. Daniel J. Figdore, Susan Ashrafzadeh-Kian, Vanessa K. Pazdernik, Alicia Algeciras-Schimnich and Joshua A. Bornhorst. Determination of pediatric and adult reference intervals for neurofilament light chain (NfL) in blood and a comparison to other recent studies J Lab Precis Med. 2024;9:29. doi:10.21037/jlpm-24-33

Method Description
Describes how the test is performed and provides a method-specific reference

Plasma calibrator or specimen are added to particle solution. Neurofilament Light (NfL) in specimens or calibrators specifically binds to anti-NfL monoclonal antibody (mouse) on the particles and antigen-antibody immunocomplexes are formed. The particles are washed and rinsed to remove unbound materials. Alkaline phosphatase (ALP)-labeled anti-NfL monoclonal antibodies (mouse) are added and specifically bind to the prior formed immunocomplexes on the particles, and additional immunocomplexes are formed. The particles are washed and rinsed to remove unbound materials. Substrate solution is added and mixed with the particles and 3-(2'-spiroadamantane)-4-methoxy-4-(3"-phosphoryloxy) phenyl-1,2-dioxetane disodium salt (AMPPD) contained in the substrate solution is dephosphorylated by the catalysis of ALP indirectly conjugated to particles. Luminescence (at a maximum wavelength of 477 nm) is generated by the cleavage reaction of dephosphorylated AMPPD. The luminescent signal reflects the amount of NfL in the sample.(Unpublished Mayo method)

PDF Report
Indicates whether the report includes an additional document with charts, images or other enriched information

No

Day(s) Performed
Outlines the days the test is performed. This field reflects the day that the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time before the test is performed. Some tests are listed as continuously performed, which means that assays are performed multiple times during the day.

Monday through Friday

Report Available
The interval of time (receipt of sample at Mayo Clinic Laboratories to results available) taking into account standard setup days and weekends. The first day is the time that it typically takes for a result to be available. The last day is the time it might take, accounting for any necessary repeated testing.

1 to 3 days

Specimen Retention Time
Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded

3 months

Performing Laboratory Location
Indicates the location of the laboratory that performs the test

Rochester

Fees
Several factors determine the fee charged to perform a test. Contact your U.S. or International Regional Manager for information about establishing a fee schedule or to learn more about resources to optimize test selection.

  • Authorized users can sign in to Test Prices for detailed fee information.
  • Clients without access to Test Prices can contact Customer Service 24 hours a day, seven days a week.
  • Prospective clients should contact their account representative. For assistance, contact Customer Service.

Test Classification
Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR) product.

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information
Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Clinic Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.

CPT codes are provided by the performing laboratory.

83884

LOINC® Information
Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the order and results codes of this test. LOINC values are provided by the performing laboratory.

Test Id Test Order Name Order LOINC Value
NFLP Neurofilament Light Chain, P 101281-4
Result Id Test Result Name Result LOINC Value
Applies only to results expressed in units of measure originally reported by the performing laboratory. These values do not apply to results that are converted to other units of measure.
NFLP Neurofilament Light Chain, P 101281-4

Test Setup Resources

Setup Files
Test setup information contains test file definition details to support order and result interfacing between Mayo Clinic Laboratories and your Laboratory Information System.

Excel | Pdf

Sample Reports
Normal and Abnormal sample reports are provided as references for report appearance.

Normal Reports | Abnormal Reports

SI Sample Reports
International System (SI) of Unit reports are provided for a limited number of tests. These reports are intended for international account use and are only available through MayoLINK accounts that have been defined to receive them.

SI Normal Reports | SI Abnormal Reports

Test Update Resources

Change Type Effective Date
New Test 2025-03-18