Patient Preparation:
1. Fasting preferred.
2. Do not collect a specimen for the 48 hours following a plasma exchange.
Collection Container/Tube: Light-blue top (3.2% sodium citrate)
Submission Container/Tube: Plastic vial
Specimen Volume: 1.5 mL
Collection Instructions:
1. Immediately after specimen collection, place the tube on wet ice.
2. Centrifuge between 1000-2000 x g for 10 minutes at 4 degrees C and aliquot plasma into plastic vial.
3. Freeze specimen within 30 minutes.
0.5 mL
Gross hemolysis | OK |
Gross lipemia | Reject |
Gross icterus | OK |
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Plasma Na Cit | Frozen | 14 days |
The complement system membrane attack complex (MAC) is formed by the C5b fragment along with C6, C7, C8 and several C9 molecules. This complex is recognized by multiple names, including MAC, terminal complement complex and C5b-9. Laboratory tests measure the amount of soluble C5b-9 (sC5b-9) complex. The formation of C5b-9 and sC5b-9 is a consequence of activation of the complement system by either the classical, lectin, or alternative pathways. Therefore, measurement of the sC5b-9 complex can be used as a surrogate marker of terminal complement activation via all complement pathways.
Elevated concentrations of C5b-9 are associated with the development of transplant-associated thrombotic microangiopathy (TA-TMA), a complication of hematopoietic stem cell transplant.(1-3) Patients with higher sC5b-9 concentrations at baseline may require the use of higher doses of eculizumab to treat TA-TMA,(4) especially in children. Because of this association, measurement of sC5b-9 before transplant as part of a diagnostic evaluation and then repeat measurements during therapy have been proposed as tools to follow-up patients.(5) Importantly, while the elevation of sC5b-9 has shown very high sensitivity for TA-TMA, it has shown only a modest specificity, ranging from 40% to 50%, and the increased sC5b-9 may be found in other transplant complications as well as several other conditions where complement activation may occur: immune-complex disease, infection, atypical hemolytic uremic syndrome, C3 glomerulopathies, etc. A panel of complement tests, such as AHUSD / Atypical Hemolytic Uremic Syndrome Complement Panel, Serum and Plasma, may provide additional information on the extent of the complement activation, along with the information of which pathway is most dysregulated.
< or =250 ng/mL
1. Qi J, Wang J, Chen J, Su J, et al: Plasma levels of complement activation fragments C3b and sC5b-9 significantly increased in patients with thrombotic microangiopathy after allogeneic stem cell transplantation. Ann Hematol. 2017 Nov;96(11):1849-1855
2. Horvath O, Kallay K, Csuka D, et al: Early increase in complement terminal pathway activation marker sC5b-9 Is predictive for the development of thrombotic microangiopathy after stem cell transplantation. Biol Blood Marrow Transplant. 2018 May;24(5):989-996
3. Mezo B, Horvath O, Sinkovits G, Veszeli N, Krivan G, Prohaszka Z: Validation of early increase in complement activation marker sC5b-9 as a predictive biomarker for the development of thrombotic microangiopathy after stem cell transplantation. Front Med (Lausanne). 2020 Oct 6;7:569291
4. Jodele S, Dandoy CE, Lane A, et al: Complement blockade for TA-TMA: lessons learned from a large pediatric cohort treated with eculizumab. Blood. 2020 Mar 26;135(13):1049-1057
5. Young JA, Pallas CR, Knovich MA: Transplant-associated thrombotic microangiopathy: theoretical considerations and a practical approach to an unrefined diagnosis. Bone Marrow Transplant. 2021 Aug;56(8):1805-1817
Microtiter plates are coated with monoclonal antibody specific to the C9 ring of the soluble C5b-9 (sC5b-9) complex. Controls, standards, and patient samples are exposed to the plate. After washing the plate, a horseradish peroxidase-conjugated anti-sC5b-9 complex antibody is added followed by a substrate to initiate color change.(Package insert: MicroVue SC5b-9 Plus EIA Kit. Quidel Corporation; 0980EN02, 09/2021)
Tuesday, Friday
This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.
86160
Test Id | Test Order Name | Order LOINC Value |
---|---|---|
C5B9 | SC5b-9 Complement, P | 93244-2 |
Result Id | Test Result Name |
Result LOINC Value
Applies only to results expressed in units of measure originally reported by the performing laboratory. These values do not apply to results that are converted to other units of measure.
|
---|---|---|
616921 | SC5b-9 Complement, P | 93244-2 |
Change Type | Effective Date |
---|---|
Test Status - Test Resumed | 2024-04-12 |