Detecting cobalt exposure in a random urine collection
Monitoring metallic prosthetic implant wear
This test is not useful for assessment of vitamin B12 activity.
Test Id | Reporting Name | Available Separately | Always Performed |
---|---|---|---|
COBR | Cobalt/Creat Ratio, U | No | Yes |
CRETR | Creatinine, Random, U | No | Yes |
COBR: Inductively Coupled Plasma-Mass Spectrometry (ICP-MS)
CRETR: Enzymatic Colorimetric Assay
Co (Cobalt)
Cobalt (Co)
Urine
High concentrations of gadolinium and iodine are known to interfere with most metals tests. If either gadolinium- or iodine-containing contrast media has been administered, a specimen should not be collected for 96 hours.
Supplies: Urine Tubes, 10 mL (T068)
Collection Container/Tube: Clean, plastic urine collection container
Submission Container/Tube: Plastic, 10-mL urine tube or clean, plastic aliquot container with no metal cap or glued insert
Specimen Volume: 3 mL
Collection Instructions:
1. Collect a random urine specimen.
2. See Metals Analysis Specimen Collection and Transport for complete instructions
2 mL
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Urine | Refrigerated (preferred) | 28 days | |
Frozen | 28 days | ||
Ambient | 14 days |
Detecting cobalt exposure in a random urine collection
Monitoring metallic prosthetic implant wear
This test is not useful for assessment of vitamin B12 activity.
Cobalt (Co) is rare but widely distributed in the environment. It is an essential cofactor in vitamin B12. While cobalt is an essential element, cobalt deficiency has not been reported in humans.
Cobalt is used in the manufacture of hard alloys with high melting points and resistance to oxidation. Cobalt salts are also used in the glass and pigment industry. Previously, cobalt salts were sometimes used as foam stabilizers in the brewing industry; this practice was banned due to the cardiovascular diseases it induced. The radioactive isotope of cobalt, (60)Co, is used as a gamma emitter in experimental biology, cancer therapy, and industrial radiography.
Cobalt is not highly toxic, but large doses will produce adverse clinical manifestations. Acute symptoms are pulmonary edema, allergy, nausea, vomiting, hemorrhage, and kidney failure. Chronic symptoms include pulmonary syndrome, skin disorders, and thyroid abnormalities. The inhalation of dust during machining of cobalt alloyed metals can lead to interstitial lung disease. Improperly handled (60)Co can cause radiation poisoning from exposure to gamma radiation.
Urine cobalt concentrations are likely to be increased above the reference value in patients with metallic joint prosthesis. Prosthetic devices produced by Zimmer Company and Johnson and Johnson typically are made of aluminum, vanadium, and titanium. Prosthetic devices produced by DePuy Company, Dow Corning, Howmedica, LCS, PCA, Osteonics, Richards Company, Tricon, and Whiteside typically are made of chromium, cobalt, and molybdenum. This list of products is incomplete, and these products change occasionally; see prosthesis product information for each device for composition details.
COBALT:
0-17 years: Not established
>17 years: <1.7 mcg/g Cr
CREATININE:
> or =18 years old: 16-326 mg/dL
Reference values have not been established for patients who are younger than 18 years of age.
Concentrations greater or equal to 2.0 mcg/g creatinine indicate excess exposure. There are no Occupational Safety and Health Administration blood or urine criteria for occupational exposure to cobalt.
Prosthesis wear is known to result in increased circulating concentration of metal ions. In a patient with a cobalt-based implant, modest increase (2-4 mcg/g creatinine) in urine cobalt concentration is likely to be associated with a prosthetic device in good condition. Excessive exposure is indicated when urine cobalt concentration is greater than 5 mcg/g creatinine, consistent with prosthesis wear. Urine concentrations greater than 20 mcg/g creatinine in a patient with a cobalt-based implant suggest significant prosthesis wear. Increased urine trace element concentrations in the absence of corroborating clinical information do not independently predict prosthesis wear or failure.
Specimen collection procedures for cobalt require special specimen collection tubes, rigorous attention to ultraclean specimen collection and handling procedures, and analysis in an ultraclean facility. Unless these precautions are taken, elevated urine cobalt results may be an incidental and misleading finding.
1. Keegan GM, Learmonth ID, Case CP. A systematic comparison of the actual, potential, and theoretical health effects of cobalt and chromium from industry and surgical implants. Crit Rev Toxicol. 2008;38(8):645-674
2. Lhotka C, Szekes T, Stefan I, Zhuber K, Zweymuller K: Four-year study of cobalt and chromium blood levels in patients managed with two different metal-on-metal total hip replacements. J Orthop Res. 2003;21(2):189-195
3. Lison D, De Boeck M, Verougstraete V, Kirsch-Volders M. Update on the genotoxicity and carcinogenicity of cobalt compounds. Occup Environ Med. 2001;58(10):619-625
4. Sodi R.
Vitamins and trace elements. Rifai N, Chiu RWK, Young I, eds: Tietz Textbook of Laboratory Medicine. 7th ed. Elsevier; 2023:chap 39.
5. Crutsen JRW, Koper MC, Jelsma J, et al. Prosthetic hip-associated cobalt toxicity: a systematic review of case series and case reports. EFORT Open Rev. 2022;7(3):188-199
6. Leyssens L, Vinck B, Van Der Straeten C, Wuyts F, Maes L. Cobalt toxicity in humans-A review of the potential sources and systemic health effects. Toxicology. 2017;387:43-56. doi:10.1016/j.tox.2017.05.015
The metal of interest is analyzed by inductively coupled plasma mass spectrometry.(Unpublished Mayo method)
Tuesday, Wednesday, Friday
This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.
83018
82570
Test Id | Test Order Name | Order LOINC Value |
---|---|---|
COBRU | Cobalt/Creat Ratio, Random, U | 13468-4 |
Result Id | Test Result Name |
Result LOINC Value
Applies only to results expressed in units of measure originally reported by the performing laboratory. These values do not apply to results that are converted to other units of measure.
|
---|---|---|
CRETR | Creatinine, Random, U | 2161-8 |
607272 | Cobalt/Creat Ratio, U | 13468-4 |